Circ Cardiovasc Interv. 2024 Apr 12:e013889. doi: 10.1161/CIRCINTERVENTIONS.123.013889. Online ahead of print.
ABSTRACT
BACKGROUND: Anatomic selection for Harmony valve implant is determined with the aid of a screening report and perimeter plot (PP) that depicts the perimeter-derived radius along the right ventricular outflow tract (RVOT) and projects device oversizing. The PP provides an estimation of suitability for implant, but its sensitivity as a screening method is unknown. This study was performed to describe anatomic features and outcomes in patients who underwent Harmony TPV25 implant despite a PP that predicted inadequate oversizing.
METHODS: We reviewed RVOT anatomic features and measurements in patients who underwent transcatheter pulmonary valve replacement with the Harmony TPV25 device despite a PP that predicted inadequate oversizing.
RESULTS: This study included 22 patients. There were no unsuccessful implants or adverse valve-related events. Anatomic features varied, but all patients fit into 1 of 3 anatomic types characterized by differences in RVOT dimensions. Type 1 patients (n=9) had a long RVOT with a choke point and a wide main pulmonary artery. Type 2 patients (n=6) had a short RVOT that was pyramidal in shape, with no choke point, and extensive main pulmonary artery lengthening/expansion during systole. Type 3 patients (n=7) had a short, bulbous main pulmonary artery with a choke point and an open pulmonary artery bifurcation.
CONCLUSIONS: Transcatheter pulmonary valve replacement with the Harmony valve is feasible in some patients whose PP fit analysis predicts inadequate oversizing. All cases in this series fit into 1 of 3 anatomic patterns, which are not identified in the screening report. Implanters must review cases individually to assess the feasibility of the implant.
PMID:38606564 | DOI:10.1161/CIRCINTERVENTIONS.123.013889