Blood Press. 2024 Dec;33(1):2411294. doi: 10.1080/08037051.2024.2411294. Epub 2024 Oct 11.
ABSTRACT
OBJECTIVE: Findings from adult studies suggest that tronco-conical cuffs provide more accurate blood pressure (BP) measurements in individuals with obesity. The aim of the present study was to examine differences in office blood pressure (BP) levels using conical cuffs compared to standard-shaped cylindrical cuffs in children and adolescents with obesity.
DESIGN AND METHOD: We performed an observational study, including 37 children and adolescents with obesity, who were consequently recruited from the outpatient clinics of the Obesity and Cardiovascular Risk Unit at General University Hospital Consortium of Valencia. Arm circumference AC was measured in all participants, and the appropriate cuff size was selected for both conical and cylindrical cuffs.
RESULTS: Mean participants’ age was 11.8
CONCLUSIONS: Although the use of conical cuffs in the study showed no advantage in enhancing the performance of BP measurements, they may be considered an alternative for office BP measurements in children and adolescents with obesity. Their reliability should be confirmed in larger populations and different settings.
PMID:39391937 | DOI:10.1080/08037051.2024.2411294