J Clin Lab Anal. 2024 Apr 15:e25037. doi: 10.1002/jcla.25037. Online ahead of print.
ABSTRACT
BACKGROUND: In newborns, elevated nucleated red blood cell (NRBC) levels can be associated with enhanced erythropoietic stress and might be predictive for adverse outcome. Also, the presence of NRBC in peripheral blood might lead to erroneous enumeration results of white blood cells in automated hematology analyzers. We aimed to assess the comparability of the Sysmex XN 1000 to manual slide reviews and correlation of NRBC with inflammation markers.
METHODS: Specimens of 3397 children under 1 year were compared by automated and microscopic NRBC enumeration. Additionally, potential correlations between NRBC and age and inflammation markers were examined.
RESULTS: Overall, there was good correlation (r = 0.97) between automated (range: 0%-3883%) and microscopic enumeration (range: 0%-3694%) of NRBC with high comparability up to a NRBC value of 200% and an increase in the variation between the two methods with increasing NRBC numbers. When 94 samples with ≤ 200% NRBC and ≥ 30% divergence between methods were separately reanalyzed with respect to overlapping cell populations in their scattergrams, Sysmex would have generated unrecognized incorrect automated results in 47 samples, corresponding to 1.4% of total study samples. NRBC counts were negatively correlated to age, but not to inflammation markers.
CONCLUSION: Sysmex XN 1000 is highly precise in the enumeration of NRBC in children under 1 year up to counts of 200% and might replace time-intense manual counting in routine diagnostics. In the setting of neonatal and intensive care diagnostics, microscopic control and supervision of scattergrams are highly recommended for any automated NRBC enumeration processes.
PMID:38619294 | DOI:10.1002/jcla.25037