Spine Deform. 2024 Mar 31. doi: 10.1007/s43390-024-00854-7. Online ahead of print.
ABSTRACT
PURPOSE: This study evaluates the intraoperative and short-term complications associated with robotically assisted pedicle screw placement in pediatric posterior spinal fusion (PSF) from three surgeons at two different institutions.
METHODS: We retrospectively reviewed 334 pediatric patients who underwent PSF with robotic-assisted navigation at 2 institutions over 3 years (2020-2022). Five thousand seventy robotically placed screws were evaluated. Data collection focused on intraoperative and early postoperative complications with minimum 30-day follow-up. Patients undergoing revision procedures were excluded.
RESULTS: Intraoperative complications included 1 durotomy, 6 patients with neuromonitoring alerts not related to screw placement, and 62 screws (1.2%) with documented pedicle breaches, all of which were revised at time of surgery. By quartile, pedicle breaches statistically declined from first quartile to fourth quartile (1.8% vs. 0.56%, p < 0.05). No breach was associated with neuromonitoring changes or neurological sequelae. No spinal cord or vascular injuries occurred. Seventeen postoperative complications occurred in eleven (3.3%) of patients. There were five (1.5%) patients with unplanned return to the operating room.
CONCLUSION: Robotically assisted pedicle screw placement was safely and reliably performed on pediatric spinal deformity by three surgeons across two centers, demonstrating an acceptable safety profile and low incidence of unplanned return to the operating room.
PMID:38556583 | DOI:10.1007/s43390-024-00854-7