Pediatr Emerg Care. 2024 Jul 1. doi: 10.1097/PEC.0000000000003155. Online ahead of print.
ABSTRACT
OBJECTIVES: The aim of this study was to create and validate a 3-dimensional (3D) ultrasound model with normal and abnormal pediatric hip joint anatomy that is comparable to a pediatric hip joint in appearance and anatomy and replicates sonographic characteristics of a pediatric hip joint.
METHODS: A 3D rendering of the bone and soft tissue was created from a computed tomography pelvic scan of a pediatric patient. This rendering was modified to include a unilateral joint effusion. The bone was 3D printed with a photopolymer plastic, whereas the soft tissue was cast with a silicone mixture in a 3D-printed mold. The effusion was simulated by injecting saline into the soft tissue cavity surrounding the bone. The ultrasound model was validated by pediatric point-of-care ultrasonographers at an international pediatric ultrasound conference.
RESULTS: A pediatric hip ultrasound model was developed that simulates both normal and abnormal pediatric hip joint anatomy, each with an appropriately sized, measurable joint effusion. Validation by pediatric point-of-care ultrasonographers showed that the key aspects of a normal pediatric hip joint (femoral physis, sloped femoral neck, and adequate soft tissue) with an identifiable and measurable effusion were included in the ultrasound model.
CONCLUSIONS: In this study, we successfully created a cost-effective, reusable, and reproducible 3D pediatric hip ultrasound model. The majority of pediatric point-of-care ultrasonographers who evaluated the model agreed that this model is comparable to a pediatric patient for the purpose of teaching ultrasound skills and joint space measurement.
PMID:38950383 | DOI:10.1097/PEC.0000000000003155